Home arrow Laboratory Equipment arrow Class II Safety Cabinets arrow BioAir S@feMate
  • German
header_bg.gif
 

Upcoming events

 
BioAir S@feMate Print E-mail
 

Microbiological Safety Cabinet 

    Fully EN12469 certified by Tüv
     Hamburg
    State of the art microprocessor
     control system
    Large digital display, high resolution
    Air and aerosol-tight sliding sash,
     electrically operated by finger touch
    Alarms for low air flow and wrong
     front window position
    Sloped front and back wall for the
     most comfortable access
    Front access for filter maintenance
     and service
    C-shaped support stand for the
     easiest one man installation
     procedure
    Easy retrofit option kits

 
  Article No.: 14297 (Model 1.2) & 14303 
     
(Model 1.5)
 

 
Safemate Cabinets are supplied in three different sizes (1.2 mt and 1.5 mt). These last generation Microbiological Safety Cabinets Class II typ A2, have been certified according to the most stringent safety standards (EN 12469-2000).
The internal design, the air flow aerodynamics and monitoring, the built-in safety devices and the very accurate manufacturing, guarantees the highest performances at the most stringent safety levels, as specified by EN 12496 standard and have been certified by the most prestigious European certification bodies for Safety Cabinets. Certified intrinsic biological safety, combined with impressively competitive prices, give the end user a state of the art cabinet accessible to every budget, that only experienced European design and accurate quality manufacturing can provide.
 
    Microprocessor controlled motor blower,
     with volumetric sensor for exhausted air
     flow monitoring
    State of the art Microprocessor control
     system offering:
      Large screen monitor
      Automatic control of preset airflow
       volumes
      Sliding sash window with smart control
      Permanent monitoring of HEPA filters
       life span
      Alarms. Multilevel alarms, with
       redundancy functions
      Permanent display of working
       conditions
      Highest air flow stability both in case of
       transitional disturbances or to
       progressive filter clogging
      Semi-automatic fumigation cycle
       (EN12297 tested and certified)
      Continuous monitoring of front barrier
       air flow for the highest operator safety
      Low barrier alarm
      Power failure alarm
    Volt-free contact for remote monitoring
     of exhaust fan.
    Automatic reset of initial conditions in
     case of power failure
    C-shaped support stand for the easiest
     one man installation procedure

   
    Sloped front design for the highest
     operational comfort. Sloped back side of
     the working chamber for the best down
     flow distribution (cabinet carcass
     EN12298 tested and certified for air
     tightness)
    Utilities inlets from the top of the
     cabinet.
    Stainless Steel internal surfaces with 2B
     finishing (including spillage tray). Solid
     work surface (3 sections) and special
     designed front grill.
    Electrically operated sliding multilayer
     safety glass window
     (max opening at 120°)
    Comfortable 200mm front opening
    Easy to install retrofit options through
     lateral sides.
    Exposed exhaust Hepa filter for easy
     visual integrity check.
    H14 class High Efficiency Particulate Air
     filters with 99.999% efficiency
     on .3micron particles (most penetrating
     particle diameter) (EN1822-1 and EN
     13091:1999 tested and certified)
    Both exhaust and Main Filters are
     equipped with a micromesh membrane
     located downstream which acts as
     airspeed equalizer expansion plenum, as
     well as a clear indicator of filter
     damages.
    Filter change and maintenance from the
     front of the cabinet.
    Exhaust transitions easily installable.
    Key operated. The key can be removed
     when the unit is in SAFE mode, in order
     to avoid unwanted operation. In case of
     power failure, the cabinet is re-set to
     original working conditions.

   
    Self calibration cycle performed when
     cabinet is switched on.
    High speed rinse and set up cycle
     performed, before reaching the SAFE
     operating mode.
    Visual display of SAFE conditions. Pre-
     warning before actual alarm conditions
     are reached (visual and acoustic alarms)
    Soft touch control with keys for standard
     service utilities. Interconnected UV and
     fluorescent lights.
    Exhaust and recirculating flow rates
     ensure 25 air changes/min in the working
     area (30% 70% split)
    Front barrier air speed ≥ 0.5mt/sec
    Aperture protection Factor (Apf) ≥ 1.5 x
     10 exp 5
    Cleanability Index CC grade.(EN 12296
     tested and certified)
    Light intensity on work surface
     > 900 lux.
    Noise level ≤ 56dB(A) 1.2 mt Model
     (ISO 11201, ISO 4871 and ISO EN 3744
     tested and certified)
    Work surface displacement (vibration)
     <0.005mm RMS between 20Hz and
     20,000Hz (ISO 5349 tested and certified)
    230V, 50Hz
    Max power (for each power point)
     3Amps.
    Microprocessor equipped with analogical
     watch dog.
    CE certification according to Machinery
     Directive 89/392/EEC, 91/368/EEC,
     93/44/EEC 93/68/EEC.

 