InControl 1050 - Technical details |
|
|
| |
Technical Data |
(ambient temperature 15° - 32° C) |
|
| |
Case: |
• Plastic grey 600 g |
|
| |
CO2 measurement: |
• Range
0…10% • Resolution 0.1% • Accuracy
0…6%
+/- 0,2%
6…10%
+/- 0,3% • Gas flow during measurement 0,4 l/min • Duration of a measurement approx. 2 min. • Variable measuring interval between 2 min. and 60 min.
• Dual beam infra-red sensor |
|
| |
Temperature measurement: |
• Range
0…50° C
• Resolution 0.1° C
• Accuracy 20…50° C
+/- 0,2° C
• High quality PT1000 sensor
|
|
| |
Data logger: |
• Interval for data logging
0.25 h / 0.5 h / 0.75 h / 1 h
• Max. number of measurements that can be stored 1000 |
|
| |
Power supply |
• Rechargeable batteries
• Charger for power supply and automatic charging of the
rechargeable batteries
• Power supply of the charger 230 V~ (115 V~) |
|
| |
Operation time: |
• Rechargeable batteries
up to 3 h
• Charger
continuous operation
• Time to recharge batteries
approx. 4 h |
|
|