InControl 1050 - Technical details |
| |
Technical Data |
(ambient temperature 15 - 32°C) |
|
| |
Case: |
• Plastic housing (ABS) |
|
| |
CO2 -Measurement: |
• Measurement range 0...10%
• Display resolution 0.1%
• Measurement accuracy 0....6% +/- 0.2%
6...10% +/- 0.3%
• Sample volume during measurement 0,8 l
• Duration of a measurement approx. 2 min.
• Dual beam IR sensor |
|
| |
Temperature measurement: |
• Measurement range 0...100° C
• Display resolution 0.1°C
• Measurement accuracy 20....50°C +/- 0.2°C
50...100°C +/- 0.3°C
• High quality PT1000 sensor
|
|
| |
Data logger: |
• Continuous measurement collection in 5-minutes intervalls: 15 - 120 min.
• Max. number of measurements that can be stored: 1008 |
|
| |
Power supply |
• Rechargeable batteries (Li-Ion)
• Power supply AC 100 - 240 V, DC 5V 1,5A |
|
| |
Operation time: |
• Battery operation time approx. 4 hrs
• Battery charging time approx. 4 hrs |
|
|